Product Category

UL Horizontal Vertical Flame Chamber

UL1581 Horizontal Vertical Flame Chamber Product details:Model ZLT-HV4: The ZLT Horizontal-Vertical Flame Chamber Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. the device complies with UL1581-2001(1060 FT1,1061,1080VW-1,1090,1100FT2)…

UL1581 Horizontal Vertical Flame Chamber
 Product details:Model ZLT-HV4: 
 The ZLT Horizontal-Vertical Flame Chamber Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. the device complies with UL1581-2001(1060 FT1,1061,1080VW-1,1090,1100FT2) standards-related requirements designed. Applicable to the single wire and cable, and soft-line horizontal (FT1), vertical (VW1) combustion test. Burning time and burning the number of adjustable, can be automatic / manual, vertical / horizontal switching, gas pressure, flow adjustable and equipped with flame temperature recognition device.

SPECIFICATIONS

Flame standard

ANSI/ASTM D 5207 & UL 1581

Burner Tube

ANSI/ASTM D 5025

Burning Angle

0°,20°

Thermocouple

Type K (Ni/Cr – Ni/Al),200°~1050°

Thermocouple Size O/D

0.5mm

Temperature Resolution

1oC

Standard Copper Block

weight:10 ± 0 .05 g before drilling. Cu-ETP


Time for Temperature to

Increase

1、225W/50mm(16.5mm)/84±2S  
     A:405 ±10mL/min& 45±5 mm water
     2、500W/125mm(40mm)/54±2S  
     A:965 ±30mL/min&125±25 mm water
     B:385 ±15mL/min&552±100 mm water
     C:300 ±10mL/ min&600±125 mm water 

Gas Supply

methane,min.98% purity

Input Power

220 V 50 Hz 3A or 115 V 60HZ  Optional

Volume Of Chamber

>4m3, black interior

Outer Dimensions

W*D*H=2000mm*1000mm*2000mm

Conform to standard

UL1581:1060/1061/1080/1090/1100

Includes: Flame height gauge and 100% absorbent cotton

Order info:
 Product Type:  
 Place of origin: Guangzhou ,China
 Price Terms: FOB, CIF ,EXW
 Payment Terms: T/T  
 Package: wooden case
 Minimum Order: 1
 Delivery Time: 20 working days
 OEM: Yes,

Inquiry